Fluke 87 III Bedienungsanleitung Seite 28

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 58
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 27
80 Series III
Users Manual
20
To estimate capacitance values above 5 µF, use the
current supplied by the meter’s resistance function,
as follows:
1. Set up the meter to measure resistance.
2. Press Kto select a range based on the
value of capacitance you expect to measure
(refer to Table 6.)
3. Discharge the capacitor.
4. Place the meter’s leads across the capacitor;
then time how long it takes for the display to
reach OL.
5. Multiply the charge time from step 4 by the
appropriate value in the µF/second of Charge
Time column in 6. The result is the estimated
capacitance value in microfarads (µF).
Table 6. Estimating Capacitance Values Over
5 Microfarads
Expected Capacitance
Suggested
Range*
µF/second
of Charge
Time
Up to 10 µF4 Me 0.3
11 µF to 100 µF 400 ke 3
101 µF to 1000 µF 40 ke 30
1001 µF to 10,000 µF4 ke 300
10,000 µF to 100,000 µF 400 e 3000
*These ranges keep the full-charge time between 3.7 seconds
and 33.3 seconds for the expected capacitance values. If the
capacitor charges too quickly for you to time, select the next
higher resistance range.
Seitenansicht 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 57 58

Kommentare zu diesen Handbüchern

Keine Kommentare