Documenting Process Calibrator
Specifications
97
Temperature, Thermocouples (cont)
MEASURE DEGREES C SOURCE DEGREES C
TYPES (°C) 1 YEAR 2 YEAR 1 YEAR 2 YEAR
R
-20 to 0 2.3 2.8 1.2 1.8
0 to 100 1.5 2.2 1.1 1.7
100 to 1767 1.0 1.5 0.9 1.4
S
-20 to 0 2.3 2.8 1.2 1.8
0 to 200 1.5 2.1 1.1 1.7
200 to 1400 0.9 1.4 0.9 1.4
1400 to 1767 1.1 1.7 1.0 1.5
C
0 to 800 0.6 0.9 0.6 0.9
800 to 1200 0.8 1.2 0.7 1.0
1200 to 1800 1.1 1.6 0.9 1.4
1800 to 2316 2.0 3.0 1.3 2.0
Sensor inaccuracies not included.
Accuracy with external cold junction; for internal junction add 0.2°C
Resolution: 0.1°C
Temperature Scale: ITS-90 or IPTS-68, selectable
Compensation: ITS-90 per NIST Monograph 175 for B,R,S,E,J,K,N,T; IPTS-68 per IEC 584-1 for B,R,S,E,J,K,T;
IPTS-68 per DIN 43710 for L,U.
Temperature Coefficient: 0.05°C/°C in the range -10 to 18°C and 28 to 50°C
Common Mode Error: 0.01°C/(Common Mode Volt)